- 1.0 OBJECTIVE:To lay down a Standard Operating Procedure for Calibration of UV spectrophotometer.2.0 SCOPE:This procedure shall apply for Calibration of UV spectrophotometer in quality control departments
- .3.0 RESPONSIBILITY:
- All concerned personnel shall responsible to follow the procedure mention in this SOP.
- Concerned Department Heads shall be responsible for compliance of the procedure.
4.0 ACCOUNTABILITY:
QC and QA Head shall be accountable that this SOP is implemented as per laid down procedure.
5.0 ATTACHMENTS:
Observation and Data recording Tables for UV Calibration – Attachment – I
6.0 PROCEDURE:
General Procedure for UV Spectrophotometer
- All calibration standards used for the calibration should prepared as per template / SOP.
- Only use calibration standard which is procured from authentic source with certification and it should
- NIST traceable or Pharmacopoeial reference standards or From instrument manufacturer.
- Alternatively suitable commercial certified filters (NIST traceable) may be used.
Internal Calibration:
- Internal calibration should perform for the parameters mentioned in Attachment for Internal calibration observation sheet.
- Perform the internal Calibration as per manufacturer’s instruction.
- After completion of internal calibration the results obtained shall fill in Attachment -Internal calibration observation sheet, the format may differ base on manufacturer’s internal calibration requirement) for further evaluation and conclusion.
Note : Perform the internal calibration of UV Spectrophotometer, if option available in software
- 7.0 REFERENCES:ISO 2859-1:19998.0 ABBREVIATIONS:SOP: Standard Operation Procedure9.0 DISTRIBBUTION LIST:Quality Assurance
Human Resource and administration
Production
Quality Control
Warehouse
Engineering
Purchase
10.0 HISTORY OF REVISION:
Version No. Effective Date Reason for revision