SOP of Media Decontamination and Disposal after used
- Objective
To lay down the procedure for Decontamination and Disposal of Used Media
- Scope
This SOP is applicable for Decontamination and Disposal of Used Media. (Pharmaceutical Company Name).
- Responsibility
- Microbiologist or above of Microbiology Laboratory.
- Head – Microbiology section / Nominee.
- Accountability
- Head – Quality Control. / Nominee.
- Abbreviations and Definitions
SOP : Standard operating procedure
No. : Number
QC : Quality Control
QA : Quality Assurance
- Procedure
- Wear nose mask and disposable hand gloves during scrapping and decontamination activity of used media.
- For Solidified Media(Media Used for Settle plates exposure, Surface Monitoring test plates, Microbial limit test plates, Growth promotion test media, Sub-culture slants etc.):
- Segregate and arrange the plates showing growth of colonies into biohazard bag.
- Scrap out the media from the plates having no growth with the help of spatula into another biohazard bag.
- Place the bags in the Vertical autoclave.
- For Liquefied Media: (Sterility Tubes/canisters, Positive Control tubes, culture tubes product enrichment tubes etc.)
- Place all the test tubes in the test tube stand and arrange them in the Vertical autoclave.
- Place the Disposable media plates/strips/media kits, used cotton plugs, gloves used
- for handling of microbial cultures, lyophilized culture glass pieces, sterility test vent filters & tubing etc in a biohazard bag and load in the vertical autoclave.
- Keep one or two deodorizing pearls (if available) in the decontamination load to prevent putrefied smell of the used media.
- Operate the vertical autoclave as per SOP No
- After sterilization, collect the media contained in tubes/ bottles and other decontaminated material into biohazard bag.
- Add 50mL of 20% Virosil solution or 2.5% Savlon to the bag containing decontaminated media and shake the bag for proper dispersion of the added disinfectant solution.
- Discard the contents of the bag into the washbasin/sink.
- Open the tap and allow the water to flow while discarding the contents.
Forms and Records (Annexures)
-
- Not applicable
- Distribution
- Master copy – Quality Assurance
- Controlled copies – Quality Assurance, Production, Quality Control, Stores, Engineering and Human Resource
- History:
Date | Revision Number |
Reason for Revision |
00 | NEW SOP |
For More Pharma Updates Visit –https://pharmaguidances.com