SOP of LAF Operation, Calibration, Cleaning, and Maintenance
- Objective
To lay down the procedure for Operation, Calibration, Cleaning and Maintenance of LAF.
- Scope
This SOP is applicable for Operation, Calibration, Cleaning and Maintenance of LAF in Microbiology Laboratory of (Pharmaceutical Company Name).
- Responsibility
- Microbiologist or above of Microbiology Laboratory: Preparation of SOP.
- Head – Microbiology section / Nominee: Checking of the SOP.
- Accountability
- Head – Quality Control / Nominee: Compliance of SOP.
- Abbreviations and Definitions
SOP : Standard operating procedure
No. : Number
QC : Quality Control
QA : Quality Assurance
LAF : Laminar Air Flow
- Procedure
Make: Clean Room Presentations, Model No.: CRP HL(S) 5000 & 4000
- Operation:
- Connect the LAF to the Mains Power supply and switch ON the mains.
- The front panel of the LAF consists of the following buttons
- GERMICIDAL LIGHT
- Press ‘ON’ button of UNIDIRECTIONAL FLOW to turn ON the blower.
- Press ‘ON’ button of FLUORESCENT LIGHT to turn ON the Light.
- Press ‘ON’ button of GERMICIDAL LIGHT to turn ON the GERMICIDAL LIGHT.
- Before starting the work, Switch on the Blower and GERMICIDAL LIGHT for 30 minutes. After 30 minutes switch off the GERMICIDAL LIGHT and Switch the FLOUORESCENT LIGHT on. Then slowly lift the front door up, to open.
- Monitor the Positive pressure reading on Magnehelic gauge. It should be between 10 and 20 mm of water gauge.
- Wipe the LAF Work bench with 70% IPA and then start the work as per requirement.
- After completion of work, monitor the pressure reading on Magnehelic gauge.
- Clean the work station with 70% isopropyl alcohol.
- Close the door.
- Press ‘OFF’ button of FLUORESCENT LIGHT to turn OFF the Light.
- Press ‘OFF’ button of UNIDIRECTIONAL FLOW to turn OFF the Blower.
- Press ON button of GERMICIDAL LIGHT to turn ON the GERMICIDAL LIGHT and keep it for 10 minutes.
- Then Press OFF button of GERMICIDAL LIGHT to turn OFF the GERMICIDAL LIGHT.
- Record the LAF usage details in the format given as Annexure No.
- Calibration:
- Perform the calibration of Magnehelic gauge once in a year by external party.
- Cleaning and Maintenance:
- Clean the inner walls, grill and workbench of LAF Cabinet before commencement and after completion of each work with the help of lint free cloth wetted with disinfectant solution.
- Clean the outer body of the LAF using a lint free cloth wetted with disinfectant solution twice (at the start and end of day’s work).
- For Cleaning of the pre-filter remove the pre-filter from the LAF and then clean it as per SOP No.:
- Pre-filter shall be cleaned by maintenance technician once in a month with compressed air and soap water.
- Place the cleaned Pre-filter at its position. Switch on the LAF unit. Check the positive pressure on Magnehelic gauge. It should be between10 and 20 mm of water gauge.
- If pressure reading on Magnehelic gauge is not achieved, put a status label `Under Maintenance’ & ask the Maintenance Department for necessary action.
- If the reading is above 20 mm of water gauge, change the HEPA filter of the LAF.
- Forms and Records (Annexures)
- LAF Log Book
- Distribution
- Master copy – Quality Assurance
- Controlled copies – Quality Assurance, Production, Quality Control, Stores, Engineering and Human Resource