Operation, cleaning & calibration of pH meter
1.0 PURPOSE:
To lay down the procedure for operation & calibration of pH meter.
2.0. SCOPE :
This SOP shall be applicable for operation & calibration of pH meter of SVP manufacturing facility
Equipment Name | Make & Model No. | Equipment ID |
pH meter |
3.0 REFERENCE(S) & ATTACHMENT(S):
3.1 References
3.1.1 In-house
- Attachments
3.2.1 Attachment- I : Calibration record of pH meter
4.0. RESPONSIBILITY:
4.1 Operator
4.1.1 To perform the operation as per SOP.
4.2 Production person.
4.2.1 To ensure implementation of SOP.
4.3 Production Head
4.3.1 To ensure implementation of the defined system.
5.0 DISTRIBUTION
- Quality Assurance
- Production
6.0 DEFINITION & ABBREVIATION(S)
6.1 Definitions
N/A
6.2 Abbreviation
SOP : Standard Operating Procedure
QA : Quality Assurance
PROD. : Production
C.C.No. : Change Control Number
KCL : Potassium Chloride
HCL : Hydrochloric Acid
WFI : Water For Injection
SVP : Small Volume Parenteral
7.0 PROCEDURE :
7.1 Operation:
7.1.1 Ensure the cleanliness of instrument before using it.
7.1.2 Ensure that instrument is located in clean and dry place.
7.1.3 Connect the power cable of instrument with main power supply.
7.1.4 Switch ON the main power supply.
7.1.5 Switch “ON” the instrument & wait for 10 second for warming up the instrument.
7.1.6 Ensure that the instrument is calibrated. If not, calibrate the instrument first before starting the operation.
7.1.7 Wash the electrode with distilled water and wipe gently with tissue paper.
7.1.8 Take the sample solution in glass beaker & allow to attain the temperature up 25°C ±2° C or the specified temperature.
7.1.9 Dip the electrode in the test solution which pH to measured. Ensure that bulb of the electrode is properly dipped in solution.
7.1.8 Keep the instrument in “MEAS” mode & instrument will measure the temperature & pH of the same solution.
7.1.9 Allow the solution to stabilize the pH with the electrode & Note the stable reading displayed on the digital display.
7.1.10 Lift the electrode carefully and wipe gently with tissue paper.
7.1.11 Always immerse the electrode in the 3 M KCL solution, if not in use.
7.2 KEY FUNCTIONS
7.2.1 CAL : To perform CALIBRATION within the selected operating Mode.
7.2.2 HOLD : To freeze the measured reading. Press again to resume live reading.
7.2.3 MODE : Toggles between available measurement modes, pH/Temp, mV/Temp.
7.2.4 ENTER: To acknowledge the entered value.
7.2.5 MEAS : To measure the pH & temperature.
7.2.6 MI : MI (memory insert) store values in memory.
7.2.7 MR : MR (memory recall) recall values from memory.
7.3 Cleaning:
7.3.1 Ensure that the Instrument is in “OFF” position.
7.3.2 Wipe the body of pH meter with a clean wet, cloth and finally wipe with dry clean, dry cloth.
7.3.3 Remove any dirty spot present on the body of Instrument with the help of Colin cleaner.
7.3.4 After every pH measurement, wash electrode with purified water and wipe with soft tissue paper before dipping in any liquid under test.
7.4 Maintenance:
7.4.1 Keep the electrode immersed in WFI.
7.4.2 Prevent the cell to dry, if it is dry; soak for 5 to 10 minutes or longer in WFI or 1 M KCl solution.
7.4.3 Do not strike the cell against any hard surface.
7.5 Instructions for storage and Reactivation of pH meter Electrode:
7.5.1 Whenever required, fill the electrode with saturated solution of “KCl” below the filling hole.
7.5.2 Never touch the lower sensitive bulb of the electrode by hand or any other hard surface.
7.5.3 For reactivating the electrode, dip electrode in 0.1 N HCl for overnight whenever required.
7.5.4 Use electrode very carefully.
7.5.5 Whenever required change the electrolyte of electrode and KCL solution in which the electrode remains dipped.
7.5.6 Always keep electrode in 3M KCL when pH meter is not in use.
7.6 Calibration:
7.6.1 Rinse the electrode with WFI & remove the excess water with tissue paper & clean the temperature sensor also with WFI.
7.6.2 Put the instrument in “CAL” mode & press “CAL” key of the instrument.
7.6.3 Instrument will demand for pH 7.0 buffer.
7.6.4 Dip the electrode & temperature sensor in prepared first buffer solution A (7.0) pH value & allow to stabilize the measuring value of the buffer.
7.6.5 As the displayed measured value get stable press enter & instrument will demand for either 4.0 pH buffer or 10 pH buffer.
7.6.6 Rinse the electrode with WFI & soak the excess water with tissue paper.
7.6.7 Dip the electrode in buffer (4.0) pH value allow to stabilize the measuring value of the buffer.
7.6.8 After stabilization of second buffer, press enter, remove the buffer & rinse the electrode with WFI & soak the excess water with tissue paper.
7.6.9 Dip the electrode in third buffer (10.0) pH value & after stabilization the same buffer press enter.
7.6.10 Wash the electrode with WFI & press the “MEAS” to verify the calibration.
7.6.11 Mention the observed pH value data of calibration in the calibration record as per attachment : I.
7.6.12 Calibration Frequency : Daily before use.
7.6.13 Use the buffer solution readily prepared in QC on weekly basis.
8.0 REVISION HISTORY
Version No. | 00 | Effective date | |
Detail of revision : New SOP |
Attachment-I
Company logo | Company Name & address | |
Calibration Record of pH meter | Page No. x of y | |
Date | Time | Buffer solution (4.0) | Buffer solution (7.0) | Buffer solution (10.0) | Checked by (prod.) | Verified By (QA) |
Observed value
(limit : 4.0±0.05) |
Observed value
(7.0±0.05) |
Observed pH
(10.0±0.05) |
||||
Frequency: Daily before use