TABLE OF CONTENT
Description | Page No |
Performance Qualification Protocol Approval | |
Objective | |
Scope | |
Responsibility | |
Pre requisite | |
Reference | |
Testing Procedure | |
· Airflow Laminarity by Smoke Test | |
· Non viable particle count | |
· Viable particle count | |
Performance qualification summary and conclusion | |
Abbreviation | |
Performance Qualification Approval | |
Particle count test report sheet | |
Viable particle count test report sheet |
Protocol Preparation and Approval
Preparation and Checking:
Approval:
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Objective:
The objective of the protocol is to establish documentary evidence for the Open Fronted Containment Facility to demonstrate that unit is qualified for correct operation as per guidelines outlined in this protocol. Scope: This procedure applies to the performance qualification of Open Fronted Containment Facility comprising of the following tests 1. Air flow laminarity by smoke test 2. Non viable particulate count. 3. Viable particle count Responsibility: Client: 1. Preparation, Review and approval of the protocol. 2. Providing required support during operation and preparation of PQ report. 3. Execution of Viable Particle count Contractor(TEKNOPAK): 1. Execution of protocol. 2. Collection of data and preparation of final PQ report. Pre Requisite:
References: United State Pharmacopoeia 39 ISO 14644 TESTING PROCEDURE AIR FLOW LAMINARITY TEST BY SMOKE TEST Purpose: – The purpose of this test is to check the laminarity of Air flow which is coming through HEPA filters of the unit. Apparatus Required: – Rod with cotton wrapped on one end. Video camera for live demonstration of the activity. Chemical required Titanium Tetra chloride : B. N —————- Acceptance Criteria:- The smoke generated by the titanium tetrachloride should flow in a laminar way when the unit is in working condition. Operating Procedure:- Switch on the unit, one hour before taking the measurement and allow the system to stabilize. Assure that the magnehlic gauge reading is with in range (10 – 20 mm). Wrap the rod one end with cotton and dip it in the Titanium tetra chloride solution. Now move the rod inside the OFCF area from left to right selecting three locations (upper, middle and lower). Cover the activity by video film. Note: Follow safety precautions, avoiding the splashing of the reagent as it fumes densely. Attachment: – 1. Snaps of video film 2. Video film disk NON VIABLE PARTICULATE COUNT Purpose: – The purpose of this test is to verify the cleanliness ISO Class-5 of the Installation in accordance with ISO 14644. Apparatus Required: – Discrete particulate counter. Acceptance Criteria:-
Operating Procedure: Set up the particle counter in accordance with the manufacturer’s instructions. Place the discrete particle counter in the sample locations. Position of sample probe pointing in to the airflow. Sample the volume of air. Record the results in the test record sheet. Sampling points 1 2 3 4 5 Attachment Calibration certificate of Discrete particle counter. Results Sheet VIABLE PARTICLE COUNT Purpose: – The purpose of this test is to demonstrate the capability of the installations air handling system to maintain the viable particle count as per USP-39 through settle plate technique and air sampler technique. Apparatus Required: – Settling Plate. Air Sampler Acceptance Criteria:- Settling Plate: Colony forming Unit = Less than 1cfu per 90 mm Petri plate. Air Sampler: Colony forming Unit = Less than 3cfu per cubic meter of air. Operating Procedure:- Settle Plate technique: Switch on the unit, one hour before taking the measurement and allow the system to stabilize. Assure that the magnehelic gauge reading is with in range (10 – 20 mm). Keep the settling plate at the ‘3’ location as defined in the drawing. Expose the plate for ‘1’ hour. Analyze the plate for Microbiological analysis. Air Sampler technique: Switch on the unit, one hour before taking the measurement and allow the system to stabilize. Assure that the magnehelic gauge reading is with in range (10 – 20 mm). Start the Air sampler as per its operational manual Sample the air at the center of the LAF safe zone by setting air sampling rate at 6m3/hr and sampling time =10 min. Take the filter and analyze the plate for Microbiological analysis Note: Sampling shall be taken for ‘3’ consecutive days Sample Location for Settle plate: 1 2 3 Attachment: Results Sheet. Performance Qualification Summary and conclusion: _________________________________________________________________ Abbreviation
Performance Qualification Approval The performance qualification documents is studied and approved by the Undersigned Authorized Personnel. The approved Performance Qualification supersedes all the previous agreed specification. Report Preparation and Review:
Report Approval:
|
Mr. Shiv Kumar is the Author and founder of pharmaceutical guidance, he is a pharmaceutical Professional from India having more than 14 years of rich experience in pharmaceutical field.
During his career, he work in quality assurance department with multinational company’s i.e Zydus Cadila Ltd, Unichem Laboratories Ltd, Indoco remedies Ltd, Panacea Biotec Ltd, Nectar life Science Ltd. During his experience, he face may regulatory Audit i.e. USFDA, MHRA, ANVISA, MCC, TGA, EU –GMP, WHO –Geneva, ISO 9001-2008 and many ROW Regularities Audit i.e.Uganda,Kenya, Tanzania, Zimbabwe. He is currently leading a regulatory pharmaceutical company as a head Quality. You can join him by Email, Facebook, Google+, Twitter and YouTube