Functioning , Calibration, Cleaning and Maintenance of LAF
Objective:
- Technique for functioning, Calibration, Cleaning and Maintenance of LAF
Scope:
- This SOP is relevant for Operation, Calibration, Cleaning and Maintenance of LAF in Microbiology Laboratory.
Responsibility:
- Chemist or above of Microbiology Laboratory.
- Head – Microbiology Section.
Accountability:
- Head – Quality Control.
Procedure:
Make: Clean Room Presentations
Operation:
- Plug in the LAF to the Mains Power supply and switch ON the mains.
- The front panel of the LAF includes the subsequent buttons
- UNIDIRECTIONAL FLOW
- FLUORESCENT LIGHT
- GERMICIDAL LIGHT
- Power ‘ ON’ button of UNIDIRECTIONAL FLOW to turn ON the blower.
- Power ‘ON’ button of FLUORESCENT LIGHT to turn ON the Light.
- Power ‘ON’ button of GERMICIDAL LIGHT to turn ON the GERMICIDAL LIGHT.
- Before Starting the work, Switch on the Blower and GERMICIDAL LIGHT for half-hour. After half-hour transfer off the GERMICIDAL LIGHT and Switch the FLOUORESCENT LIGHT on. Then slowly uplift the the front door up, to open.
- Monitor the Positive pressure analyzing on Magnehelic gauge. It have to be among 10 and 20 mm of water gauge.
- Wipe the LAF Work bench with 70% IPA after the work as per requirement.
- After completion of work , screen the pressure analyzing on Magnehelic gauge.
- Clean the work station with 70% IPA
- Close the door.
- Power ‘OFF’ button of FLUORESCENT LIGHT to turn OFF the Light.
- Power ‘OFF’ button of UNIDIRECTIONAL FLOW to turn OFF the Blower.
- Power ON button of GERMICIDAL LIGHT to turn ON the GERMICIDAL LIGHT and keep it for 10 minutes.
- Then Power OFF button of GERMICIDAL LIGHT to turn OFF the GERMICIDAL LIGHT.
- Record the LAF usage in the format given as Annexure No. A01
Calibration:
- Perform the calibration of Magnehelic gauge once in a year through external agencies
Cleaning and Maintenance:
- Clean the internal walls, grill and workbench of LAF Cabinet before beginning and after completion of every work with the assist of lint free cloth dipped with disinfectant solution.
- Clean the outer frame of the LAF the usage of a lint free cloth dipped with disinfectant solution twice (at the start and end of day’s work).
- Remove the pre-filter from the LAF for Cleaning purpose.
- Pre-filter shall be cleaned by maintenance person once in a month with compressed air and soap water.
- Place the wiped clean Pre-filter at its position. Switch on the LAF unit. Check the positive pressure on Magnehelic gauge. It should be in the middle of 10 and 20 mm of water gauge.
- If pressure analyzing on Magnehelic gauge is not achieved, put a status label `Under Maintenance’ & ask the Maintenance Department for necessary action.
- Change the HEPA filter of the LAF if the reading is above 20 mm of water gauge
For Other Pharma Related post and Updates Click Here
List of Annexure / Formats:
S.
No. |
Format Title | Annexure
Number |
1 | LAF Log Book | A01 |
References (if any):
- Not applicable.
Reason for Revision:
- Periodic review
Abbreviations:
- SOP : Standard operating procedure
- NO: Number
- QC : Quality Control
- QA : Quality Assurance
- LAF : Laminar Air Flow